Foxboro IDP10-AF1C01F I/A Series Differential Pressure Transmitter
Description
Foxboro IDP10-AF1C01F Product Description
Product Overview
Basic Information
Manufacturer: Foxboro (Invensys, now Schneider Electric)
Model: IDP10-AF1C01F
Product Type: Intelligent Differential Pressure Transmitter, Two-Wire Process Measurement Device
Applicable System: Foxboro I/A Series DCS; compatible with HART-enabled control systems, PLCs, and third-party industrial devices
Mounting Style: Direct process mounting; optional bracket kits for surface, duct, or manifold mounting
Protection Rating: IP66/IP67 (NEMA 4X) for housing; dual seal certified to ANSI/ISA 12.27.01-2003
Certifications: CE, UL, CSA, ATEX, IECEx compliant; certified for hazardous area installations (Ex ia/Ex d); meets ISA S71.04 G3 standards
Sensor Type: Piezo-resistive silicon strain gauge
Wetted Materials: 316L stainless steel (standard); optional Hastelloy C, Monel, Tantalum for corrosive media
Power Supply: 24 VDC (12–42 VDC wide range); two-wire loop-powered
Operating Temperature: -40°C to +85°C (process); -20°C to +60°C (ambient)
Storage Temperature: -50°C to +100°C
Relative Humidity: 0% to 100% non-condensing
Weight: Approximately 1.2 kg
Dimensions: 120 mm (H) × 80 mm (W) × 60 mm (D)
Country of Origin: United States
Warranty: Standard 5-year manufacturer warranty
Function and Positioning
Core Performance Parameters
Measurement Range: Differential pressure from -30 bar to +30 bar (-435 psi to +435 psi); standard spans include 75 mbar, 500 mbar, 2.1 bar, 21 bar, 210 bar
Accuracy: ±0.1% of calibrated span (standard); ±0.075% optional for high-precision applications
Repeatability: ±0.01% of calibrated span
Stability: ±0.05% of span per year (long-term drift)
Output Signal: 4–20 mA DC analog (loop-powered); superimposed HART 7 digital communication (1200 baud)
Response Time: ≤100 ms (standard); adjustable damping from 0 to 60 seconds
Static Pressure Rating: Up to 250 bar (3625 psi) for high-pressure applications
Temperature Compensation: Automatic compensation for ambient and process temperature variations
Electrical Isolation: 1500 VAC dielectric isolation between process and electrical circuits
Diagnostics: On-board LCD with pushbuttons for local configuration; LED status indicators for power, communication, and faults; remote diagnostic support via HART
Vibration Resistance: Compliant with ISA S71.04 G3; withstands 5–200 Hz vibration at 0.15 mm amplitude
General Adaptable Accessories
Mounting Hardware: Surface mounting brackets, manifold mounting kits, pipe clamps, vibration isolation mounts, stainless steel fasteners
Communication Accessories: HART communicators, HART modems, USB-to-HART adapters, configuration software (Foxboro Toolkit, HART Field Communicator)
Process Connections: 1/4” NPT, 1/2” NPT, DIN 19213 flanges; remote seal assemblies with capillary tubes; diaphragm seals for high-temperature or viscous media
Power Supplies: 24 VDC loop power supplies, redundant power modules, surge protectors, intrinsic safety barriers for hazardous areas
Replacement Parts: Sensor modules, LCD displays, pushbutton assemblies, housing gaskets, terminal blocks, wiring kits
Protection Kits: Weatherproof enclosures, sun shields, insulation jackets for extreme temperatures, EMI shielding gaskets
Calibration Tools: Pressure calibration pumps, precision pressure gauges, HART calibration software, loop simulators